How your 3499A/B/C Test System
Benefits From Upgrading to the New
34980A Switch/Measure/Control System
Application Note
Add functionality, streamline your test
programs, and reduce the physical test
system size by upgrading from your
3499A/B/C to the new 34980A
34980A Mainframe, plug-in modules and terminal cards
Multifunction Switch/Measure Unit.
This application note will describe
differences and advantages of the
new 34980A as compared to the
3499A/B/C. The 34980A
Switch/Measure System masters
switching, measurement, and
control features by offering a
superior selection of switch
topologies and system control
features. A state-of-the-art DMM
has been integrated into the
34980A mainframe providing a
wide variety of measure capability.
The 34980A mainframe also
includes standard PC connections
and instrument control options for
your ease of use.
measurements are acquired. The
34980A also has configurable
Measurements
One of the advantages of the new
34980A is its ability to make
highly accurate test system
measurements. The integration of
a high-speed DMM into the
mainframe enhances the 34980A
switch and control system by
providing voltage, current,
temperature, frequency, and
resistance measurements without
using any of the mainframe’s
available eight slots. The DMM
conveniently generates
High/Low limits on each channel
that activate an alarm if the input
signals are out of the set range.
The 34980A’s integrated DMM
saves you time and money by
reducing the number of individual
instruments required, and
additional cabling for
measurements and triggering. In
comparison, the 3499A/B/C
switch requires an external DMM
or other measurement instrument
as well as additional wiring to
provide the same functionality.
measurement results in selectable
engineering units. Eleven
different types of measurements
can be made and are listed on the
page 2. The 34980A time-stamped
measurement data can be stored
or displayed and analyzed as the
Discover the detailed advantages of
upgrading your test system using
the 34980A in this document.
multiplexer switch module, a
counter/timer and digital I/O with
pattern generator/detector along
with digital I/O and D/A
converters. The 34980A RF and
microwave switch drivers offer
functionality for a broader range
of RF/microwave switches being
used by customers today.
The 3499A/B/C does provide a
selection of optics switching
which is not available with the
34980A. Table 2 shows the
34980A and 3499A/B/C system
modules organized by switch
and/or measurement type. It is
easy to compare and upgrade
from the 3499A/B/C switch
modules to the new 34980A
switch modules as well as add
new switch, measurement and
control capabilities.
Modules
Benefits of the new 34980A
modules vs 3499A/B/C modules
The 34980A implements a
superior selection of switch
topologies. The 34980A switch
topologies are similar to
3499A/B/C switch modules and
include multiplexers, matrixes,
high-power, general purpose, and
RF/microwave switches.
However, the 34980A modules
have utilized the newest switch
technology which enables a higher
density of channels per module,
and the ability to achieve the
faster scan rates with the
integrated DMM. The 34980A
offers other functionality used in
test systems and includes digital
I/O, a counter/totalizer and D/A
converters. New 34980A
•
Increased channel count per
module with the same switch
reliability
•
Increased drivability for
various RF/uWave switches
•
•
•
•
•
High Speed FET multiplexer
Reed Matrix switch
Switch/attenuator driver
Counter/Timer
High Speed Digital I/O with
memory
•
•
Analog output with waveform
capabilities
Breadboard module
with +5V, +12V
functions include a FET
With so many different types of
switch modules to choose from,
you can easily identify the
modules from Table 2 that
provide the best solution to meet
your test application needs. For
detailed module specifications,
please refer to the product data
sheets available at these web
locations:
http://www.agilent.com/find/34980A
and
3499A/B Mainframe and plug-in modules
3
Table 2. Direct comparison of 34980A and 3499 modules
Module Type
34980A
3499A/B/C
Multiplexer
34921A 40 Ch armature 300V 1A
34922A 70 Ch armature 300V 1A
34923A 40 Ch reed 200V .5A
34924A 70 Ch reed 80V .5
34925A 40 Ch FET +/-80V .5A
34937A 28 Ch, 1A and 4 Ch, 5A
34938A 20 Ch, 5A
N2260A 40 Ch armature 200V 1A
N2266A 40 Ch reed 200V .5A
N2270A 10 Ch 1000V 1A
44470A 10 Ch armature 250V 2A
44470D 20 Ch armature 250V 2A
N2261A 40 Ch
N2267A 8 Ch, 8A
44471A 10 Ch
44471D 20 Ch
General Purpose
Matrix
44477A 7 Ch SPDT (Form C)
N2262A 4x8 armature 200V 1A
44473A 4x4 armature 250V 2A
34931A Dual 4x8 armature 300V 1A
34932A Dual 4x16 armature300V 1A
34933A Dual 4x8 reed 80V .5A
Digital I/O
See multifunction 34950A, 34952A
N2263A 32-bit TTL
44474A 16-bit TTL
Multifunction
34952A 32-bit Digital I/O with 2 Ch
DAC and Counter/Timer
34950A 64-bit Digital I/O with Pattern
Generator/Detector
N2264A 12 GP, 3 GP 5A, 16-bit Dig I/O
N2265A 4x4 matrix, 16-bit Dig I/O
N2269A 2 DAC, 16-bit Dig I/O
Fiber-Optic Multiplexer
RF & Microwave
N2280A Quad 1x2
N2281A Dual 1x4
N2282A Single 1x8
34941A Quad 1x4 RF 1.5G 50Ω, SMA
34946A 4G/20G 50Ω, 2 SPDT SMA
(N1810TL)
34947A 4G/20G 50Ω, 3 SPDT, SMA
(N1810UL)
N2268A Dual 1x4 3.5G 50Ω
N2272A Single 1x9 1.0G 50Ω
N2276A Dual 1x6 20G 50Ω
(87106B/4B)
44472A Dual 1x4 300M 50Ω
44478A Dual 1x4 1.3G 50Ω
44476A Triple 1x2 18G 50Ω
(8762B)
34942A Quad 1x4 RF 1.5G 75Ω, mSMB
34945A uwave Switch/Attn Driver
32 SPDT or 8 attenuators
The following µwave switches and
attenuators are supported with
the distribution1 boards;
44478B Dual 1x4 1.3G 75Ω
N2276B Relay driver (2 switches)
Choose from switches;
87104A/B/C
87106A/B/C
also drive two external attenuators
84904/6/7/K/L
N181x series SPDT switches
876X series SPDT switches
87104x/106x multi-port switches
87606x series matrix switches
8722x transfer switches
44476B Relay driver (2 switches)
Choose from switches;
8762A/B/C/F
849x series attenuators
8763B/C
8490x series attenuators
8764B/C
Generic screw terminal connections
Third-party switches (screw terminals)
34951A Quad 16-bit D/A Converter
34959A +12 V, +5 V
DAC
Breadboard
See multifunction
44475A 5.5
4
User Interface
connecting to Agilent instruments
you can access the Agilent
similar to the 3499A/B/C’s but
they are not interchangeable. A
comparison of the commands is
provided in appendix A. The
34980A's IVI-C and IVI-COM
drivers are similar to Plug & Play
drivers, the 3499A/B/C's Plug &
Play driver for example, but
include improvements. The IVI
(Interchangeable Virtual
The 34980A offers the most
modern instrument I/O options for
interfacing a controller PC with the
switch system: GPIB, USB and
LAN. All three Input/Output
interface connections come
Technologies USB/LAN/GPIB
Interfaces Connectivity Guide at
With LAN, the instrument can be
located closer to the device under
test without worrying about the
distance to the PC. A USB to
instrument connection is even
easier. The computer
standard with the 34980A
mainframe. If you choose to modify
or upgrade your test system I/O,
the 34980A is system ready.
Instruments) driver specification
was built upon the Plug&Play
driver specification but adds the
capabilities provided by the
automatically detects instruments
that are USB ready. If you are
currently using the 3499A/B/C
You can also easily control the
34980A without connecting to a PC
by pressing the front-panel buttons,
an advantage when first setting up
or troubleshooting a system.
RS232 interface, it is recommended Microsoft ActiveX technologies
to upgrade to the 34980A GPIB,
USB or LAN interface to achieve
optimal data transfer rates.
which simplify instrument
interchangeability, and product
performance. The 34980A also
includes many new commands for
the operation of the embedded
DMM. A 34980A Web interface
provides a new easy way to
communicate via LAN with the
instrument.
Upgrading your test system from
using the 3499A/B/C GPIB to the
34980A GPIB, or newer USB or
LAN is quick and easy. If you have
used I/O Config from the Agilent
I/O Libraries to set your GPIB
instrument connection, you would
use the same I/O Config to set up
your LAN or “TCPIP” instrument
connection. Retrieve the LAN,
TCPIP address from your
instrument and enter it into I/O
Config. When you address your
instrument in a program, it will
look similar to; TCPIP0::9::INSTR.
For more information on
Instrument Control
(Programming)
The 34980A and 3499A/B/C can
both be controlled locally from the
front panel, or by sending either
instrument or driver commands
from a PC for remote control. The
34980A’s instrument commands
and drivers are based on modern
standards such as SCPI (standard
commands for programmable
instruments) and IVI-C and IVI
COM drivers. The SCPI instrument
commands of the 34980A are
Table 3 shows a summary of the
programming control options
available for the 3499A/B/C and the
34980A. There is similarity between
the instrument commands, however,
the 34980A has evolved to use the
most modern and widely accepted
instrument command standards.
Table 3. Control Options for 3499A/B/C and 34980A
3499A/B/C
34980A
Front Panel Buttons
3488 Legacy Commands
SCPI Commands
Plug & Play driver
IVI COM driver
G
G
G
G
G
G
G
G
G
G
G
IVI-C driver
LabVIEW driver
Web Interface
5
Listed below are each of the many ways available to communicate and
control the 34980A, the communication options and their description
Web Page Interface
The 34980A comes with a built-in
detailed instrument web page that
allows you to remotely connect and
communicate with the mainframe
and installed modules. You
establish a LAN connection, either
directly between the 34980A and
PC, or from the 34980A to an
existing Local Area Network. The
instrument Web page resides on
the 34980A and can be accessed
using a standard web browser such
as Netscape or Internet Explorer
on a PC. Just type the IP address
into the web browser. The web
interface allows you to select any of
the installed modules and view the
current configuration, change
settings, make measurements, and
makes it easy to troubleshoot or
debug the switch/measure system.
The instrument web interface also
allows you to check the current
status of the instrument by viewing
the error queue, calibration status
and the accumulated module
Web interface display for 34921A MUX Switch module
switch closures.
Web interface display of scanned measurements
6
SCPI Commands
The 34980A can be programmed
using SCPI commands or the IEEE
standard commands.
Configuration, measurements and
system control of the mainframe
and modules are easily achieved
using these instrument commands.
and channel(s) rather than only
three digits as the 3499A/B/C
uses. Also, for the 3499A/B/C the
[ROUTe] in the command
structure is optional, where with
the 34980A it is required.
IVI-C, IVI COM and LabVIEW
instrument drivers
34980A IVI-C, IVI COM and
LabVIEW instrument drivers are
available for simplified
programming. If you currently
program using Plug&Play drivers,
the IVI-C driver is similar and can
be used to program in the same
way. The IVI COM driver conforms
to the most modern IVI COM
instrument driver standard and
can be used in any programming
environment. LabVIEW
The 34980A offers new
measurement and module
capabilities such as counter/timer
One difference between
instrument commands for the
3499A/B/C and the 34980A relates and digital I/O with pattern
to addressing specific channels in
the system. Table 4 shows an
example difference in 3499A/B/C
switch control commands
compared to 34980A commands.
Notice the 34980A uses four digits
to address the module slot number
generator/detector that use
commands unique to their
functionality. Appendix A includes
a list of the 34980A and 3499A/B/C
commands as well as direct
programmers are able to use the
34980A LabVIEW instrument
driver for their graphical
comparisons of the commands.
programming environment.
Table 4. Command comparisons
34980A
3499A/B/C
Close channel 10 on the module in slot 3
ROUTe:CLOS (@3010)
[ROUTe:]CLOS (@310)
Scan channels 0 through 40 on the module in slot 1
ROUTe:SCAN (@1001:1040)
[ROUTe:]SCAN (@100:140)
Example program. 34980A SCPI commands in VEE
7
Although the 34980A scan feature
can be used with either the
The 3499A/B/C cannot store
readings but does offer more scan
configurability. For example,
once a 3499A/B/C scan list is set
up, the user can select an arm
source, a trigger source, the
number of sweeps, and the delay
time for each individual channel
within the scan list.
Scan Feature
The scan feature is a sequence of
alternating switch closures and
measurements. This feature is
provided by both the 34980A and
a 3499A/B/C combined with a
DMM or other measurement
instrument. The 34980A increases
the measurement scan abilities
and offers more than twice the
number of scanned channels
within one scan list compared to
the 3499A/B/C as seen here.
internal DMM or with an external
instrument, using the integrated
DMM provides a faster and more
efficient scan rate. The integrated
DMM eliminates the need for
external instrument triggers and
wiring. The 3499A/B/C scan list
requires the use of instrument
triggers to work with an external
DMM or other measurement
instrument. Each switch module is
rated with individual switch scan
rates. A comparison of the scan
rates for both systems Multiplexer
and Matrix switch modules is
shown in Table 5 below.
Single Scan List Size:
34980A
Up to 560
2-wire channels
or
Another advantage the 34980A
system has over the 3499A/B/C
system is its ability to store
Up to 640
1-wire channels
measurement readings. When the
34980A internal DMM is used, up
to 500,000 time stamped readings
can be stored in volatile memory.
3499A/B/C Up to 200 channels
and/or bits
Table 5. 34980A and 3499A/B/C Multiplexer & Matrix module scan rates
Module Type
34980A
Modules
Scan Rate
Ch/s
3499A/B/C
Modules
Scan Rate
Ch/s
Multiplexer
34921A 40 Ch armature
34922A 70 Ch armature
34923A 40 Ch reed
100
100
500
500
1000
100
100
500
N2260A 40 Ch armature
N2266A 40 Ch reed
N2270A 10 Ch
80
350
100
43
34924A 70 Ch reed
44470A 10 Ch armature
44470D 20 Ch armature
N2262A 4x8 armature
44473A 4x4 armature
34925A 40 Ch FET
43
Matrix
34931A Dual 4x8 armature
34932A Dual 4x16 armature
34933A Dual 4x8 reed
80
43
8
The 3499A/B/C monitor feature
allows users to continuously
monitor a selected switch or
module status from the front
display. A channel monitor can
be set up using the front panel
buttons or instrument commands.
The 3499A/B/C monitor output
can be a specific switching
channel, a digital I/O port or the
state of all switches or digital I/O
on one plug-in module. The
3499A/B/C allows a single
channel to be monitored
Monitor Feature
Stored State Feature
The 34980A monitor feature
allows you to monitor readings
acquired by the switch/measure
system. You can also monitor
readings acquired with the
34980A internal DMM. To
monitor a measurement from the
front panel, simply select the
channel and press either the
Measure [DMM] or [Channel]
button to see a continually
updated channel output or state
on the front display. For
The 34980A is also able to
store/recall up to six (0 through
5) instrument setups. A stored
setup state can be automatically
recalled at any time, even during
a power up state. The stored
state feature contains the state of
the plug-in modules including
channel configurations, scanning
setups, alarm values, and Mx + B
scaling values. Location 0 is
typically used for the Factory
Reset and power down state, but
can also be programmed with a
new setup.
continuously even during an
instrument scan.
programs, use the
ROUTe:MONitor:DATA?
Command. The command reads
the data from the channel
The 3499A/B/C is able to store
and recall up to 50 instrument
setup states with revision 4.0.
The instrument setups include the
status of relay channels, and/or
the static digital I/O state, module
configuration, and scanning
setups (scan lists, arm count, arm
source, etc.). The 3499A/B/C is
not able to automatically power
up in a stored state.
currently selected by the Monitor
function. One reading is displayed
for each monitor command. The
Web interface can also be used to
monitor a channel measurement.
9
(Dsub50 and Dsub78) are low cost
and readily available from many
sources. The detachable screw
terminal blocks are easy to wire
and include built-in cable strain
relief. The screw terminals attach
to the switch modules with a
secure latching mechanism for
ease of use and improved
connection reliability. No tools
are required to easily connect and
disconnect the wired terminals.
The 3499 terminals are secured to
The 34980A terminal wiring and
cables are different from that of
the 3499A/B/C. However, they
offer a better solution to test
system interface due to the use of
highly reliable standard
Switch and Measurement
Connections
Most often, test systems must be
flexible enough to be easily moved
to a new location or modified for
different devices under test.
connectors and low cost cables.
Having a removable wiring
terminal block or a pre-wired cable
is optimal in these situations. The
module’s terminal and/or cable
can simply be removed and
replaced with a different terminal
for the new test configuration. This
prevents the need to rewire the
same terminal each time for
Refer to Table 2 for the modules
that have optional terminal
blocks or the 34980A data sheet
for specific module wiring
connection details:
the module and housing by screws http://www.agilent.com/find/34980A.
at the end of the terminals and
require a screwdriver to remove
or replace.
The 3499A/B/C offers removable
screw terminal blocks for many of
the modules and also an option for
pre-wired DIN96 to quad D25, or
dual D50 cables. Some of the
modules such as the Optics,
Microwave and RF modules
include the SMA or BNC terminals
for application specific
different applications.
Fortunately, both the 34980A and
3499A/B/C provide the removable
terminals and pre-wired cables so
changing your test system wiring
from the 3499A/B/C to the 34980A
is fairly simple.
For more complex test system
wiring, mass interconnect (one
location in a system where the
switch, instruments and DUT are
interconnected) is made easy by
using solutions provided by
The 34980A offers the ability to
use either standard Dsub 50 and
Dsub 78 connectors/cables,
detachable terminal cards or mass
interconnect connections. The RF
and Microwave modules include
standard connectors such as SMA
and SMB. The 34980A cables
connection experts. The standard
Dsub 50 and Dsub 78 connectors
make it faster and easier for Mass
Interconnect suppliers to provide a
test system wiring solution. The
easy to use connections facilitate
time and cost savings for test
systems using the 34980A.
connections. Refer to Table 2 for
modules that include terminals.
34980A Modules and Std DSub 50/78 connectors
3499A/B/C Modules and Terminals
10
There are a few specialized
features the 3499 offers that are
not available with the new
34980A. It is good to be aware of
the differences as you upgrade to
the newer system.
Features that differentiate
new 34980A:
Summary
Although the 3499A/B/C
•
FET switches for long life, high
speed measurement scans and the
latest in technology Armature and
Reed relays for both Multiplexer
and Matrix switching
Switch/Control system has been a
great addition to many test
systems, the new 34980A now
offers an even better module
selection plus measurement
capability, standard PC
connections for communication,
and is easy to connect to modules.
The wide offering of plug-in
modules for the 34980A enables
you to select a configuration for
your specific application needs.
3499A/B/C differentiating features
•
•
Analog bus for joining modules,
connecting to external instruments,
accessing internal DMM.
•
4 dig I/O ports built into mainframe
doesn’t require a module
•
•
Fiber-Optic modules
Up to 5A general purpose switch
addresses the most common high
power and device control
applications.
1000V multiplexer module
The 34980A Switch/Measure unit
can be used in the same
applications as the 3499A/B/C
including design verification,
functional test and data
•
•
Bread board with +12V and +5V
supplies for more design flexibility.
Integrated DMM and measurement
capability save you time versus
programming with an external
measurement instrument, and
money with the switch and DMM
in one instrument.
acquisition applications. It can
also be used in many new
applications such as for data
logging, as a data acquisition
system and/or as a switch system.
It can even provide a solution for
applications requiring transducer-
based measurements such as
thermocouple or strain. The
34980A offers a great switch,
measurement and test system
control solution at a great value.
•
•
Up to 1000 readings/sec scan rate
for higher speed applications
GPIB and new USB and LAN
interfaces provide flexibility, lower
cost, higher data transfer rates,
and are the emerging trend of
choice of interfaces.
•
•
Up to 70 Ch per module offer
higher density solutions
Standard module connectors &
cables offer an easy connection
and for Mass interconnect
solutions.
•
Screw terminals with latching
mechanisms are easy to remove
and replace and securely fastened
when in use.
•
•
Web interface to quickly and easily
test or debug the test system
switching.
Large memory buffer for acquired
measurements.
11
Appendix A – Command Reference/Comparison
34980A Commands
3499A/B/C Commands
ABORt
ABORt
ARM
:SOURce
BUS|EXTernal|IMMediate|TIMer|MIX/HOLD
:SOURce?
:COUNt
<number>|MIN| MAX
:TIMer
<seconds>|MIN|MAX
:TIMer?
CALCulate
:AVERage:AVERage?
:AVERage:CLEar
:AVERage:COUNt?
:AVERage: MAXimum?
:AVERage:MAXimum:TIME?
:AVERage:MINimum?
:AVERage:MINimum:TIME?
:AVERage:PTPeak?
:LIMit:LOWer
:LIMit:LOWer?
:LIMit:LOWer:STATe
:LIMit:LOWer:STATe?
:LIMit:UPPer
:LIMit:UPPer?
:LIMit:UPPer:STATe
:LIMit:UPPer:STATe?
:SCALe:GAIN
:SCALe: GAIN?
:SCALe:OFFSet
:SCALe:OFFSet?
:SCALe: OFFSet:NULL
:SCALe: STATe
:SCALe: STATe?
:SCALe:UNIT
:SCALe: UNIT?
CALibration
:COUNt?
:LFRequency?
:SECure:CODE
SECure:STATe
:SECure:STATe?
:SETup
:SETup?
:STRing
:STRing?
:VALue
:VALue?
CALibration?
CONFigure
CONFigure
:EXTernal[:TRIGger]
:SOURce
:SOURce?
[:OUTPut]
0|1|OFF|ON
[:OUTPut]?
12
34980A Commands (cont’d.)
3499A/B/C Commands (cont’d.)
:CURRent:AC
:CURRent[:DC]
:DIGital:STATe
:DIGital:WIDTh
:FREQuency
:FRESistance
:PERiod
:RESistance
:TEMPerature
:TOTalize
[:VOLTage]:AC
[:VOLTage][:DC]
CONFigure?
DATA
:LAST?
:POINts:EVENt:THReshold
:POINts:EVENt:THReshold?
:POINts?
:REMove?
DIAGnostic
:DMM:CYCLes:CLEar
DIAGnostic
:DMM:CYCles?
:RELay CYCles:CLEar
:RELay:CYCLes?
[:RELay]:CYCLes:CLEar
[:RELay]:CYCLes?
[:RELay]:CYCLes:CLEar
[:RELay]:CYCLes?
[:RELay]:CYCLes:MAX?
:DISPlay[:INFOrmation]
:DISPlay:STATe?
:MONitor
:MONitor?
:SPEEK?
:SPOKE
DISPlay
[:STATe]
[:STATe]?
:TEXT
:TEXT ?
:TEXT:CLEar
FETCh?
FORmat
FETCh?
:BORDer
:READing:ALARm
:READing:ALARm?
:READing:CHANnel
:READing:CHANnel?
:READing:TIME
:READing:TIME?
:READing:TIME:TYPE
:READing:TIME:TYPE?
:READing:UNIT
:READing:UNIT?
INITiate
INITiate
INSTrument
:DMM:CONNect
:DMM:DISConnect
:DMM:INSTalled?
:DMM[:STATe]
:DMM[:STATe]?
MEASure
13
34980A Commands (cont’d.)
3499A/B/C Commands (cont’d.)
:CURRent:AC?
:CURRent[:DC]?
:DIGital[:<width>]?
:DIGital[:BYTE]:BIT?
:FREQuency?
FRESistance?
:PERiod?
:RESistance?
:TEMPerature?
:TOTalize?
[:VOLTage]:AC?
[:VOLTage][:DC]?
MEMory
:STATe:CATalog?
:STATe:DELete
:STATe:NAME
:STATe:NAME?
:STATe:RECall:AUTO
:STATe:RECall:AUTO?
:STATe:RECall:SELect
:STATe:RECall:SELect?
:STATe:VALid?
OUTPut
:ALARm<n>:CLEar
:ALARm:MODE
:ALARm:MODE?
:ALARm:SLOPe
:ALARm: SLOPe?
:ALARm<n>:SOURce
:ALARm<n>:SOURce?
[:STATe]
R?
READ?
ROUTe
[ROUTe:]
:CHANnel:ADVance:SOURce
:CHANnel:ADVance:SOURce?
:CHANnel:DELay
[CHANnel:]DELay
[CHANnel:]DELay?
:CHANnel:DELay?
:CHANnel:DELay:AUTO
:CHANnel:DELay:AUTO?
:CHANnel:FWIRe
:CHANnel:FWIRe?
:CHANnel:LABel:CLEar:MODule
:CHANnel:LABel[:DEFine]
:CHANnel:VERify[:STATe]
:CLOSe
CLOSe
:CLOSe?
CLOSe?
:MODule:BUSY?
:MODule:WAIT
:MONitor[:CHANnel]
:MONitor[:CHANnel]ENABle
:MONitor:DATA?
:MONitor:MODE
:MONitor:STATe
:MONitor:STATe?
:OPEN
OPEN
:OPEN?
OPEN?
:OPEN:ABUS
:OPEN:ALL
:OPERation:OVERlap[ENABle]
14
34980A Commands (cont’d.)
3499A/B/C Commands (cont’d.)
:SCAN
:SCAN:ADD
:SCAN:REMove
:SCAN:SIZE?
SCAN:SIZE?
SCAN[:LIST]?
SCAN[:LIST]
SCAN CLEar
CPAir
CPAir?
FUNCtion
FUNCtion?
SAMPle
SENSe
:COUNt
:COUNt?
SENSe
:DIGital
:DATA:BIT?
:DATA[:<BYTE|WORD|LWORD>[:VALue]?
:DATA[:<BYTE|WORD|LWORD>:BLOCK?
:DATA[:<BYTE|WORD|LWORD>:TRACE
:TRACe[:DATA]?
:CURRent:AC
:BANDwidth
:BANDwidth?
:RANGe
:RANGe?
:RANGe:AUTO
:RANGe:AUTO?
:CURRent[:DC]
:APERture:ENABled?
:APERture
:NPLC
:NPLC?
:RANGe
:RANGe?
:RANGe:AUTO
:RANGe:AUTO?
:RESolution
:RESolution?
:ZERO:AUTO
:DIGital:DATA[:<width>]?
:DIGital:DATA[BYTE]:BIT?
:FREQuency
:APERture
:APERture?
:RANGe:LOWer
:RANGe:LOWer?
:VOLTage:RANGe
:VOLTage:RANGe?
:VOLTage:RANGe:AUTO
:VOLTage:RANGe:AUTO?
:FRESistance
:APERture
:APERture:ENABled?
:NPLC
:NPLC?
:OCOMpensated
:OCOMpensated?
:RANGe
:RANGe?
:RANGe:AUTO
15
34980A Commands (cont’d.)
3499A/B/C Commands (cont’d.)
:RANGe:AUTO?
:RESolution
:RESolution?
:ZERO:AUTO
:FUNCtion
:FUNCtion?
:PERiod:APERture
:PERiod:VOLTage:RANGe
:VOLTage:RANGe?
:VOLTage:RANGe:AUTO
:VOLTage:RANGe:AUTO?
:PERiod:ZERO:AUTO
:RESistance:APERture
:APERture?
:ALPERture:ENABled?
:NPLC
:NPLC?
:OCOMpensated
:OCOMpensated?
:RANGe
:RANGe?
:RANGe:AUTO
:RANGe:AUTO?
:RESolution
:RESolution?
:ZERO:AUTO
:TEMPerature:ABERture
:APERture:ENABled?
:NPLC
:NPLC?
:TRANsducer:Tcouple:IMPedance:AUTo
:Tcouple:CHECk
:RJUNction
:RJUNction:TYPE
:TYPE
:TRANsducer:TYPE
:FRTD:RESistance
:FRTD:TYPE
:RTD:RESistance
:RTD:TYPE
:THERmistor:TYPE
:ZERO:AUTO
:RJUNction?
:TOTalize:CLEar:IMMediate
:DATA?
:SLOPe
:THReshold[:MODE]
:TYPE
:VOLTage
:AC:BANDwidth
:AC:BANDwidth?
:AC:RANGe
:AC:RANGe?
:AC:RANGe:AUTO
:AC:RANGe:AUTO?
[:DC]:APERture
[:DC]:APERture?
[:DC]:APERture:ENABled?
[:DC]:IMPedance:AUTO
[:DC]:NPLC
16
34980A Commands (cont’d.)
3499A/B/C Commands (cont’d.)
[:DC]:NPLC?
[:DC]:RANGe
[:DC]:RANGe?
[:DC]:RANGe:AUTO
[:DC]:RANGe:AUTO?
[:DC]:RESolution
[:DC]RESolution
[:DC]:ZERO:AUTO
[:DC]:ZERO:AUTO?
SOURce
SOURce
:CURRent[:LEVel]
:CURRent[:LEVel]?
:DIGital
:MODE
:MODE?
:CONTrol:POLarity
:CONTrol:POLarity?
:FLAG:POLarity
:FLAG:POLarity?
:IO:POLarity
:IO:POLarity?
:DATA[:<BYTE|WORD|LWORD>:POLarity
:DATA[:<BYTE|WORD|LWORD>:POLarity?
:DATA:BIT
:DATA[:<BYTE|WORD|LWORD>[:VALue]
:DATA[:<BYTE|WORD|LWORD>:BLOCK
:DATA[:<BYTE|WORD|LWORD>:TRACE
:TRACe:DEFine
:TRACe:DEFine?
:TRACe:CATalog?
:TRACe[:DATA]
:TRACe:DELete[:NAME]
:TRACe:DELete:ALL
:DIGital:DATA[:<width>]
:DIGital:DATA[BYTE]:BIT
:DIGital:STATe?
:FUNCtion:CLOCk:EXTernal:DIVisor
:CLOCk[INTernal]:PERiod
:CLOCk:SOURce
:CURRent:GAIN
:CURRent:OFFSet
:ENABle
:FREQuency
:HALT
:STATus?
:TRACe:NCYCles
:TRACe[NAME]
:TRACe:SINDex
:TRIGger:IMMediate
:TRIGger:SOURce
:VOLTage:GAIN
:VOLTage:OFFSet
:MODE
:MODE?
:VOLTage[:LEVel]
:VOLTage[:LEVel]?
STATus
STATus
:ALARm:CONDition?
:ALARm:ENABle
17
34980A Commands (cont’d.)
3499A/B/C Commands (cont’d.)
:ALARm:ENABle?
:ALARm[:EVENt]?
:MODule:ENABle
:MODule:EVENt?
:MODule:SLOT[n]:CONDition?
:MODule:SLOT[n]:ENABle
:MODule:SLOT[n][:EVENt]?
:OPERation:CONDition?
:OPERation:ENABle
:OPERation:ENABle?
:OPERation[:EVENt]?
:PRESet
:OPERation:CONDition?
:OPERation:ENABle
:OPERation: ENABle?
:OPERation[:EVENt]?
:PRESet
:QUEStionable:CONDition?
:QUEStionable:ENABle
:QUEStionable:ENABle?
:QUEStionable[:EVENt]?
SYSTem
SYSTem
:ABUS:INTerlock:SIMulate
:ALARM?
:BEEPer
:BEEPer:STATe?
:CDEScription?
:COMMunicate:ENABle
:GPIB:ADDRess
:LAN:AUTOip
:LAN:BSTatus?
:LAN:CONTrol?
:LAN:DHCP
:LAN:DNS
:LAN:DOMain
:LAN:GATEway
:LAN:HISTory?
:LAN:HISTory:CLEar
:LAN:HOSTname
:LAN:IPADdress
:LAN:KEEPalive
:LAN:MAC?
:LAN:SMASk
:LAN:TELNet:PROMpt
:LAN:TELNet:WMESsage
:CPON
:CTYPe?
:DATE
:CPON
:CTYPe?
:DATE?
:ERRor?
:ERRor?
:KLOCk:EXCLude
:KLOCk:EXCLude?
:KLOCk[:STATe]
:KLOCk[:STATe]?
:LOCK:OWNer?
:LOCK:RELease
:LOCK:REQuest?
:MODule?
:MODule:PFAil:JUMPer:AMP5?
:MODule:WIRE:MODE
:MODule:TEMPerature?
:PRESet
:SECurity:IMMediate
:TIME
18
34980A Commands (cont’d.)
3499A/B/C Commands (cont’d.)
:TIME?
:TIME:SCAN?
:VERSion?
:VERSion?
:LOCal
:REMote
:RWLock
:STATe:DELete
TRACe:CATalog?
[:DATA]
[:DATA]:DAC
[:DATA]:FUNCtion
:DELete:ALL
:DELete[:NAME]
:FREE?
:POINts?
TRIGger
TRIGger
[:IMMediate]
:COUnt
:COUNt?
:DELay
:DELay?
:DELay:AUTO
:DELay:AUTO?
:SOURce
:SOURce?
:TIMer
:SOURce
:SOURce?
:TIMer
:TIMer?
TIMer?
UNIT
:TEMPerature
:TEMPerature?
IEEE-488.2 Common Commands
*CLS
IEEE-488.2 Common Commands
*CLS
*ESE *ESE?
*ESR?
*OPC *OPC?
*IDN?
*OPC *OPC?
*PSC *PSC?
*RCL
*RST
*SAV
*SRE *SRE?
*STB?
*TRG
*TST?
*WAI
19
Get the latest information on the products and
applications you select.
Agilent T&M Software and Connectivity
Agilent’s Test and Measurement software and
connectivity products, solutions and developer
network allows you to take time out of
connecting your instruments to your computer
with tools based on PC standards, so you can
focus on your tasks, not on your connections.
more information.
Agilent Technologies’ Test and Measurement
Support, Services, and Assistance
Agilent Technologies aims to maximize the value
you receive, while minimizing your risk and
problems. We strive to ensure that you get the
test and measurement capabilities you paid for
and obtain the support you need. Our extensive
support resources and services can help you
choose the right Agilent products for your
applications and apply them successfully. Every
instrument and system we sell has a global
warranty. Support is available for at least five
years beyond the production life of the product.
Two concepts underlie Agilent’s overall support
policy: “Our Promise” and “Your Advantage.”
By internet, phone, or fax, get assistance with
all your test & measurement needs
Online assistance:
Phone or Fax
United States:
(tel) 800 829 4444
Our Promise
Canada:
(tel) 877 894 4414
(fax) 905 282 6495
Our Promise means your Agilent test and
measurement equipment will meet its advertised
performance and functionality. When you are
choosing new equipment, we will help you with
product information, including realistic
China:
(tel) 800 810 0189
(fax) 800 820 2816
performance specifications and practical
recommendations from experienced test
engineers. When you use Agilent equipment, we
can verify that it works properly, help with
product operation, and provide basic
Europe:
(tel) (31 20) 547 2323
(fax) (31 20) 547 2390
Japan:
measurement assistance for the use of specified
capabilities, at no extra cost upon request. Many
self-help tools are available.
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Korea:
(tel) (82 2) 2004 5004
(fax) (82 2) 2004 5115
Your Advantage
Your Advantage means that Agilent offers a
wide range of additional expert test and
measurement services, which you can purchase
according to your unique technical and business
needs. Solve problems efficiently and gain a
competitive edge by contracting with us for
calibration, extra-cost upgrades, out-of-warranty
repairs, and onsite education and training, as
well as design, system integration, project
management, and other professional
Latin America:
(tel) (305) 269 7500
(fax) (305) 269 7599
Taiwan:
(tel) 0800 047 866
(fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100
(fax) (65) 6836 0252
engineering services. Experienced Agilent
engineers and technicians worldwide can help
you maximize your productivity, optimize the
return on investment of your Agilent
Email: [email protected]
Product specifications and descriptions in this
document subject to change without notice.
instruments and systems, and obtain
© Agilent Technologies, Inc. 2004